VERTICAL AMPLIFIERS (CH 1 and CH 2)
- Sensitivity: 5mV/div to 5V/div, 1mV/div to 1V/div at X5
- Attenuator: 10 steps in 1-2-5 sequence. Vernier control provides full adjustment between steps
- Accuracy: ±3%, ±5% at X5
- Input Resistance: 1 MΩ ±2%
- Input Capacitance: 25 pF ±10pF
- Frequency Response: 5mV to 5V/div: DC to 30 MHz (-3dB). X5: DC to 10 MHz (-3dB)
- Rise Time: 12 ns (Overshoot ≤5%)
Operating Modes:
- CH 1: CH 1, single trace
- CH 2: CH 2, single trace
- ALT dual trace, alternating
- CHOP dual trace, chopped
- ADD agebraic sum of CH 1 + CH 2
- Polarity Reversal CH 2 only
- Maximum Input Voltage: 400 V (DC + AC peak)
SWEEP SYSTEM
- Sweep Speed: 0.1 μs/div to 2 s/div in 1-2-5 sequence, 23 steps,
- Vernier control provides fully adjustable sweep time between steps.
- Accuracy: ±3%
- Sweep Magnification: 10x
TRIGGERING
- Triggering Modes: AUTO (free run) or NORM, TV-V, TV-H
- Trigger Source: CH 1, CH 2, ALT, EXT, LINE
- Maximum External Trigger Voltage: 300 V (DC + AC peak)
- Trigger Coupling: AC 30 Hz to 30 MHz
- TV H: Used for triggering from horizontal sync pulses
- TV V: Used for triggering from vertical sync pulses
TRIGGER SENSITIVITY
- Auto Bandwidth: 100 Hz-30 MHz, Internal: 1.5 div, External: 100 mV
- Norm Bandwidth: DC to 30 MHz, Internal: 1.5 div, External: 100 mV
- TV V Bandwidth: 20 Hz-1 kHz, Internal: .5 div, External: 100 mV
- TV H Bandwidth: 1 kHz-100 kHz, Internal: .5 div, External: 100 mV
HORIZONTAL AMPLIFIER (Input through channel 2 input)
- X-Y Mode Switch selectable using X-Y switch. CH 1: X axis, CH 2: Y axis
- Sensitivity: Same as vertical channel 1
- Input Impedance: Same as vertical channel 1
- Frequency Response: DC to 1 MHz typical (-3 dB)
- X-Y Phase Difference: Approximately 3 ̊ at 50 kHz
- Maximum Input Voltage: Same as vertical channel 1
CRT
- Calibrating Voltage: 1 kHz (±10%) Positive Square Wave, 2 V p-p (±3%)